ISO4064 Standard Mechanical Water Meter Custom

KeytureMeter
ISO 4064 · Standard Mechanical Meter

ISO4064 Standard Mechanical Water Meter

A multi jet water meter designed, tested and marked to ISO 4064 and OIML R49 — Q1 to Q4 error limits, pressure and temperature classes, endurance and disturbance tests, all stated on the dial.

  • Class 2: ±5% from Q1 to Q2 and ±2% from Q2 to Q4 for water up to 30 °C
  • R100 standard, ΔP63 pressure loss class, T30 temperature class, PN16
  • DN15 – DN50, orientation H or V marked, installation class stated
  • Type-tested for pressure, endurance, disturbance and temperature

ISO 4064 / OIML R49 · Q1–Q4 · ΔP63 · T30 · PN16

Product Introduction

What Is an ISO 4064 Standard Mechanical Water Meter?

An ISO 4064 standard mechanical water meter is a conventional velocity-type meter designed, tested and marked in full accordance with ISO 4064, the international standard for meters for cold potable water and hot water. The standard is published in three parts: Part 1 sets the metrological and technical requirements, Part 2 defines the test methods and Part 3 specifies the test report format. Together they define what a water meter must do — which errors it may show at which flows, which pressures and temperatures it must withstand, how it must be marked and how it must survive endurance and disturbance tests — so that a meter bought in one country can be understood, compared and verified in another. OIML R49 is the legal-metrology counterpart, and the two documents are technically aligned.

Compliance is built into the meter’s design. The standard defines four characteristic flows: Q1, the minimum flow; Q2, the transitional flow separating the lower and upper error zones; Q3, the permanent flow at which the meter must operate continuously; and Q4, the short-term overload flow. Errors may not exceed ±5% between Q1 and Q2 and ±2% between Q2 and Q4 for water up to 30 °C, with ±3% allowed in the upper zone above 30 °C. The designer therefore chooses the measuring chamber, impeller and register so that the error curve stays inside these limits, centred well inside, over the whole range. A pressure-loss class, a temperature class, an orientation code and an installation class for flow disturbance complete the meter’s identity, all of which must be stated on the dial.

For buyers, the ISO 4064 label is a shorthand for a body of evidence. It means the meter has been type tested for accuracy, static pressure, pressure loss, flow disturbance, reverse flow, endurance in discontinuous and continuous flow, and the effect of temperature. It means each production meter is verified against the stated error limits, and that its marking can be read by an inspector anywhere. Our ISO 4064 standard mechanical meter is a multi jet design supplied in DN15 to DN50 with R100 as standard, Class 2, PN16, ΔP63 pressure loss class, T30 temperature class, orientation H or V as marked and pulse or AMR options. It is the safe, familiar choice for utilities that specify by standard rather than by brand.

Key takeaway: ISO 4064 is the common language of water metering. A meter built and marked to it can be bought, tested and accepted anywhere without arguing about definitions.

Design Advantages

Why Choose an ISO 4064 Standard Meter

Six benefits of buying to the international standard rather than to a proprietary specification.

01

Common Definitions Worldwide

Q1, Q2, Q3, Q4, R, MAP, ΔP and T classes have the same meaning in every country that uses ISO 4064. Specifications and test reports can be shared without translation of concepts.

02

Defined Error Limits

The standard fixes maximum permissible errors in both flow zones. Buyers know exactly what ‘accurate’ means, and inspectors know exactly what to test.

03

Documented Type Testing

Each design is type tested for pressure, pressure loss, temperature, flow disturbance, reverse flow and endurance. The test report format in Part 3 makes results comparable between laboratories.

04

Clear Marking

The dial states Q3, R, temperature class, pressure class, orientation, direction of flow and approval marks. Anyone who reads the dial knows what the meter is for.

05

Easy Tender Evaluation

Referencing ISO 4064 in a tender defines performance in a single line. Bids can be compared against the same requirements, with test reports as evidence.

06

Foundation for Legal Metrology

National regulations and EU MID refer to ISO 4064 and OIML R49. A meter built to the standard has a short path to local approval.

Technical Data

Specifications

Standard configuration below. Pressure-loss class, temperature class, orientation and installation class can be adapted to your regulation and tender.

General Specifications
ParameterSpecification
StandardISO 4064-1:2014, ISO 4064-2:2014, ISO 4064-3:2014; OIML R49
Meter typeMulti jet, vane wheel, velocity type, dry-dial register
Nominal diameterDN15 – DN50 (1/2" – 2")
Accuracy classClass 2 (Class 1 on request for ultrasonic types)
Maximum permissible error±5% (Q1 ≤ Q < Q2); ±2% (Q2 ≤ Q ≤ Q4) for water ≤ 30 °C; ±3% above 30 °C
Range ratio RR100 standard (R80 – R160 on request)
Temperature classT30 (0.1 – 30 °C) / T50 (0.1 – 50 °C)
Maximum admissible pressureMAP 16 bar
Pressure loss classΔP63 (≤ 0.063 MPa at Q3)
Orientation codeH (horizontal) standard; V (vertical) on request
Flow profile sensitivity classU0 / D0 to U5 / D3, stated per size
Body materialBrass or cast iron
RegisterDry-dial, magnetic coupling, 5 to 6-digit odometer
ConnectionThreaded G3/4" – G2"; flange on DN40 – DN50
MarkingQ3, R, T, MAP, ΔP, orientation, flow arrow, serial number, approval marks
Pulse outputOptional reed or Hall head
ProtectionIP65 standard; IP68 on request
Size Selection

Flow Data by Size

Q4 is the overload flow, Q3 the permanent flow, Q2 the transitional flow and Q1 the minimum flow, all expressed in m³/h.

Metrological Data — R100 (ISO 4064 Class 2)
Nominal DiameterQ4 OverloadQ3 PermanentQ2 TransitionalQ1 MinimumRange RatioPressure Loss Class
DN15 (1/2")3.1252.50.040.025R100ΔP63
DN20 (3/4")540.0640.04R100ΔP63
DN25 (1")7.8756.30.10080.063R100ΔP63
DN32 (1¼")12.5100.160.1R100ΔP63
DN40 (1½")20160.2560.16R100ΔP63
DN50 (2")31.25250.40.25R100ΔP63

Values shown for R100, with Q2 = 1.6 × Q1 and Q4 = 1.25 × Q3 as defined in ISO 4064. R80 to R160 variants are available — tell us the standard edition and approvals named in your tender and we will confirm the correct marking.

Certification

Standards & Compliance

The meter is tested against the requirements of each part of ISO 4064 and the corresponding clauses of OIML R49.

ISO 4064-1:2014 Requirements

Defines metrological classes, the flow characteristics Q1–Q4, error limits, pressure and temperature classes, marking and the technical requirements for materials and construction.

ISO 4064-2:2014 Test Methods

Specifies how each requirement is tested, including test set-up, flow stability, temperature and pressure tolerances and the number of readings needed for each test point.

ISO 4064-3:2014 Test Report

Provides a standard format for the type-test report so that results are comparable and acceptable between laboratories and national authorities.

OIML R49

The legal metrology recommendation aligned with ISO 4064, used for pattern approval and as the basis for national and regional regulation.

Endurance Tests

Discontinuous and continuous flow tests at defined flows, followed by repeat accuracy tests to confirm that wear has not pushed errors outside the limits.

Flow Disturbance Tests

Meters are tested with upstream disturbances such as bends and partly closed valves to determine the installation class U and D that must be marked.

Honors and Certificates

Every certificate has witnessed our speciality.

Explore More

Get in Touch With Us

Send Message
How It Works

Technology & Design

Two things make a meter an ISO 4064 meter: the way its error curve is built into the design, and the way the standard’s tests prove it.

Designing the Error Curve to Fit the Limits

The standard defines two error zones. In the lower zone, from Q1 to just below Q2, the error may be ±5%. In the upper zone, from Q2 to Q4, it may be ±2%. The transition is set at Q2 = 1.6 × Q1. A multi jet meter’s natural error curve is a smooth hump: low flows read low because friction takes a larger share of the torque; mid flows read slightly high; and very high flows read a little low again as the impeller slips.

Designers shape the curve by choosing the number and angle of jets, the impeller diameter and blade geometry, the bearing type and the magnetic coupling strength. A final adjustment vane in the flow path lifts or lowers the middle of the curve so that it sits symmetrically inside the limits. A well-centred curve leaves margin for the drift that wear will cause in service, which is the difference between a meter that just passes and one that stays compliant for its full verification period.

  • Q2 = 1.6 × Q1 and Q4 = 1.25 × Q3 fixed by the standard
  • Jets, impeller and bearings shape the natural curve
  • Adjustment vane centres the curve inside the limits
  • Margin allowed for drift during service
±5% / ±2% Error limits in lower
and upper zones
R100 Standard range ratio
R80 to R160 available
ΔP63 Pressure loss class
≤ 0.063 MPa

How the Standard Proves It

Type testing is a comprehensive programme. The meter is first tested for accuracy across the flow range at several temperatures and pressures. It then goes through static pressure tests at 1.6 × MAP, pressure-loss measurement, flow-disturbance tests with specified upstream configurations, reverse-flow tests and endurance tests in which the meter runs thousands of cycles of starts and stops and many hours at high flow. After each stage, accuracy is re-measured to confirm that the error curve has stayed within limits.

Results are recorded in the format of ISO 4064-3. The report lists test conditions, uncertainty, results at each flow point and conclusions, so any laboratory or authority can review it. Production meters are then verified individually or by statistical sampling against the same error limits. The marking on the dial summarises this chain of evidence.

  • Accuracy across the range at several temperatures and pressures
  • Static pressure at 1.6 × MAP
  • Pressure loss, reverse flow and disturbance tests
  • Endurance in discontinuous and continuous flow
  • Standard report format for traceable results
1.6×MAP Static pressure
test level
4 parts Q1, Q2, Q3, Q4
defined and tested
IP65 Standard protection
IP68 on request
Performance Data

Head Loss, Accuracy & Dimensions

Measured performance data for the ISO 4064 standard mechanical water meter. Use these charts to confirm pressure loss class, accuracy and installation dimensions before specifying.

0 0.01 0.02 0.04 0.06 0.08 2.5 4 6.3 10 25 Head loss MPa Flow(m³/h) 15mm 20mm 25mm 32mm 40mm 50mm 2. Head Loss Curve

Figure 1 — Head loss curve (MPa vs. flow m³/h)

Head Loss Curve

The chart shows pressure loss across the meter for DN15 to DN50. ISO 4064 classifies pressure loss by the value at Q3: ΔP63 means no more than 63 kPa, and this meter meets that class at every size. The curve rises with the square of flow, so loss at Q4 is higher than at Q3.

  • ΔP63 class: ≤ 0.063 MPa at Q3
  • Loss rises with the square of flow
  • Six standard sizes DN15–DN50
+5 +2 0 -2 -5 Q1 Q2 Q3 Q4 ± 5 % ± 2 % Error% Flow 3. Accuracy Error Curve

Figure 2 — Accuracy error curve (ISO 4064 Class 2, R100, Q1–Q4)

Accuracy Error Curve

The chart shows the two error zones of ISO 4064 Class 2 as boxes: ±5% from Q1 to Q2 and ±2% from Q2 to Q4. The meter’s error curve stays inside both, centred well away from the limits, leaving margin for wear. Above 30 °C, the upper limit widens to ±3% under the standard.

  • ±5% from Q1 to Q2, ±2% from Q2 to Q4 at ≤ 30 °C
  • Curve centred inside both zones
  • Upper limit ±3% above 30 °C
3 L H B D 4. Installation Dimension

Figure 3 — Installation dimensions (L × H × B × D)

Installation Dimensions by Size
SizeL (mm)H (mm)B (mm)D (thread)
DN151659882G3/4"
DN2019010384G1"
DN25260115100G1¼"
DN32260122102G1½"
DN40300150130G2"
DN50300158132Flange

Face-to-face lengths follow ISO 4064 for interchangeability. Confirm exact values with our engineering team before installation design.

Design Comparison

Reading an ISO 4064 Dial: What Each Marking Means

The dial of a compliant meter carries a series of codes. This table explains the most important ones.

Meaning of ISO 4064 Markings
MarkingExampleMeaningWhy It Matters
Q3Q3 = 2.5Permanent flow in m³/hDefines meter size and the continuous duty rating
RR100Ratio Q3 / Q1Sets the minimum flow that is measured accurately
TT30 or T50Water temperature classShows whether hot water is allowed
MAP16 barMaximum admissible pressureUpper pressure limit for the installation
ΔPΔP63Pressure loss class at Q3Predicts pressure available after the meter
OrientationH or VInstallation position for which meter is approvedWrong orientation voids calibration
U / DU5 D3Straight pipe upstream and downstream in diametersDefines the installation conditions
Arrow→Direction of flowWrong direction gives wrong reading

Markings follow ISO 4064-1:2014 and OIML R49. Contact our engineers for help interpreting markings in a specific tender or approval document.

Premium quality

since 2002

Coming From China, Marketing To The World

Keyture Meter is located in Picturesque Jiangbei Investment Park, NINGBO, CHINA. It is a comprehensive high-tech China ISO4064 Standard Mechanical Water Meter Manufacturers and Custom ISO4064 Standard Mechanical Water Meter Factory that integrates scientific research & development, production, sales, and service. Established in 2002 as a manufacturer and exporter of water meters for more than 22 years. The factory covers an area of 40 acres (30,000 square meters), 200 staff, 100+ types of products, product capacity of 5 million per year. Its product range covers single jet, multi-jet, volumetric piston type, Woltman type, IC card intelligent water meter, remote control water meter, optoelectrical reading direct remote control water meter...

  • 0million

    Yearly production capacity

  • 0+㎡

    Production Area

  • 0People

    Works and inspectors

  • 0+ years

    Production experience

What Sets Us Apart

Excellent product with exquisite craftsmanship

Customization

We have a strong R&D team, and we can develop and produce products according to the drawings or samples the customers offered.

Cost

We are a professional manufacturer and also have our own processing machinery. So we can offer good prices and good products directly.

Quality

We have our own testing lab and advanced and complete inspection equipment, which can ensure the quality of the products.

Service

We focus on developing high-quality products for top-end markets. Our products are in line with international standards, and are mainly exported to Europe, America, Africa, Asia and other destinations around the world.

Capacity

Our annual production capacity is over 5 million units. We can meet the needs of different customers with different purchase quantities.

Shipment

We are about 35 kilometers away from Ningbo Port. It is very convenient and efficient to ship goods to any other countries.

Solution

Complete Metering Solutions

From mechanical meters to smart AMR/AMI systems — we provide end-to-end water metering solutions tailored to your project.

Complete water metering solution by Keyture Meter
Where It Works

Applications

Chosen by buyers who specify by standard, and in markets where ISO 4064 or OIML R49 forms the basis of regulation.

International TendersMulti-country procurement where the standard is the common reference.
Utilities Specifying by StandardWater authorities that define requirements by ISO 4064 class, R and temperature.
Regulated Billing MetersMarkets where legal metrology requires approved meters.
Export ProgrammesMeters supplied under local approval derived from ISO 4064 and OIML R49.
Test LaboratoriesReference models for training and benchmarking of verification rigs.
Standardised Fleet ReplacementReplacement programmes that must maintain the same specification.
Field Guidance

Installation & Maintenance

Installation Best Practices

Read the dial before installing. Confirm that orientation (H or V), direction of flow, MAP and temperature class match the application. Install the meter in the position shown and in the direction of the arrow. A meter used in a position other than the marked one is outside the approval and may show different errors.

Follow the installation class marked on the dial. U5 D3, for example, means that 5 × DN of straight pipe upstream and 3 × DN downstream are required to guarantee the stated accuracy under disturbed flow. Where installation conditions are poorer, use a flow conditioner or a longer straight length.

Flush the pipework, install a strainer if the supply carries sediment, tighten fittings with a spanner on the nut and seal the register after commissioning. Keep the type-test report and certificate with the asset record, since they form the evidence base for the meter’s use.

Read dialH / V, arrow, MAP, T
U / DStraight pipe as marked
Keep reportsTest certificate in asset file

Maintenance & Service Life

ISO 4064 does not set the service life of a meter, but the endurance tests give a basis for expecting stable performance for a stated period. Most regulations set re-verification at 6 to 12 years, and utilities typically replace residential meters at 8 to 10 years.

In-service testing is usually done by sampling: a batch of meters is removed, tested at Q1, Q2 and Q3 using the same method as in ISO 4064-2 and compared with the error limits. If the sample passes, the batch can remain in service; if not, replacement is required.

Keep records of meter serial numbers, test results and removal dates. These records help utilities understand their fleet’s ageing, choose the right replacement cycle and justify decisions to regulators.

6–12 yrsTypical regulatory
re-verification period
SamplingQ1, Q2, Q3 tests
per ISO 4064-2
RecordsSerial, results
and removal dates
Questions

Frequently Asked Questions

What is ISO 4064?

ISO 4064 is the international standard for meters for cold potable water and hot water. It defines metrological classes, error limits, test methods and test reports. Its three parts are Requirements, Test methods and Test report format.

What do Q1, Q2, Q3 and Q4 mean?

Q1 is the minimum flow, Q2 is the transitional flow that separates the lower and upper error zones, Q3 is the permanent flow at which the meter operates continuously and Q4 is the short-term overload flow. Q2 = 1.6 × Q1 and Q4 = 1.25 × Q3.

What are the error limits for Class 2?

±5% from Q1 to just below Q2, and ±2% from Q2 to Q4 for water up to 30 °C. For water above 30 °C, the upper limit widens to ±3%. Class 1 has tighter limits of ±3% and ±1%.

What does R100 mean?

R is the ratio Q3 / Q1. R100 means the minimum flow Q1 is one hundredth of permanent flow Q3. A higher R means the meter measures accurately at lower flows.

What do the pressure loss classes mean?

ΔP63 means pressure loss at Q3 must not exceed 63 kPa (0.063 MPa). Other classes are ΔP40, ΔP25 and ΔP16 for lower loss. Lower loss meters are chosen for low-pressure supply.

What is the difference between T30 and T50?

These are temperature classes. T30 means the meter is for water from 0.1 to 30 °C, and T50 allows 0.1 to 50 °C. Hot water classes such as T30/90 are used for hot water meters.

What do U and D mean on the dial?

U is the length of straight pipe upstream and D downstream, in multiples of the nominal diameter. U5 D3 means five diameters upstream and three downstream are needed to guarantee the stated accuracy.

How is ISO 4064 related to OIML R49 and the MID?

OIML R49 is the legal metrology recommendation technically aligned with ISO 4064. The EU Measuring Instruments Directive refers to them for water meters. A meter built to ISO 4064 has a straightforward route to approval under these frameworks.

Which edition of ISO 4064 applies?

The current edition is ISO 4064:2014. Older meters may be marked to ISO 4064:1993 using letter classes A, B, C and D. Please specify the edition in your tender so we can confirm marking.

Does ISO 4064 cover smart or ultrasonic meters?

Yes. The 2014 edition covers mechanical and electronic meters, including ultrasonic and electromagnetic types, with additional tests for electronic devices. The same flow characteristics and error limits apply.

How are meters tested in service?

By sampling: a number of meters are removed and tested at Q1, Q2 and Q3 on a bench, using the same method as ISO 4064-2. Results are compared with the error limits to decide whether the batch can stay in service.

Can you supply the type test report?

Yes. Type-test reports in the ISO 4064-3 format are available for each model, together with certificates for verification and approval marks required for your market. Please tell us the destination so we can include the right documents.

Tell us which standard your tender names

Send us the standard edition, class, range ratio and approvals in your specification. Our engineers will confirm the matching ISO 4064 mechanical meter and marking, with price and lead time.

Request a Quote